ASCENT Validation

Sim4Life validation

We designed test simulations to verify ASCENT’s activation thresholds. The verifications were performed by The Foundation for Research Technologies in Society (IT’IS) with the Sim4Life (Zurich, Switzerland) simulation platform. Running the following simulations required modification of the Sim4Life solver to implement the required electrical anisotropy of tissue conductivities and the boundary condition to represent the thin layer approximation used to model the perineurium.

Monofascicular rat nerve model

We validated activation thresholds for fibers seeded in a model of a rat cervical vagus nerve instrumented with a bipolar cuff electrode (Figure A and B).

_images/Picture2_CrossSection1.jpg

Figure A. Raw histology image (r.tif), segmented histology (i.tif), and scalebar (s.tif) of a rat cervical vagus nerve sample that served as inputs to define the cross-section of the nerve in the FEM.

_images/Picture3_Cuff1.jpg

Figure B. FEM of a rat cervical vagus nerve sample instrumented with a bipolar cuff electrode.

The conductivity values applied to the rat cervical FEM are provided in Table A, and the boundary conditions applied are provided in Table B.

Table A. Conductivity values for FEM of rat cervical vagus nerve. These values were also used in multifascicular nerve model and human model verifications.

Parameter Application Resistivity Ω-m
Endoneurium Within each fascicle

1.75 longitudinal

6 radial (rat and human)

12 radial (multifascicular dummy model)

Saline Cylindrical shell between the nerve and cuff 1/1.76
Platinum For both contacts 1/(9.43*106)
Silicone For electrode body 1012
Muscle Used for the surrounding "medium": Everything outside of the nerve and cuff, other than the saline layer between the nerve and cuff

1/0.35 longitudinal

1/0.086 radial

Epineurium Within the nerve around each fascicle 6.3
Encapsulation Between cuff and nerve, and immediately surrounding cuff 6.3

Table B. Boundary conditions used in FEM of rat cervical vagus nerve, multifascicular dummy nerve, and human cervical vagus nerve.

Parameter Setting
Current conservation All domains
Initial condition V=0 (all domains)
Perineurium (sides of each fascicle)

Modeled as contact impedance

1149 Ω-m *0.03* dfasc[m]

Ground (all outer boundaries of the model) V = 0
Point current source (one in each contact)

-1 mA

    1 mA

We compared thresholds for 100 5.7 µm myelinated axons (MRG model) seeded in the cross-section of the nerve in response to a single 100 µs duration monophasic rectangular pulse. The differences in thresholds between ASCENT and IT’IS model implementations was <4.2% for all fibers, demonstrating strong agreement (Figure C).

_images/Picture4.jpg

Figure C. Comparison of activation thresholds for the rat cervical vagus nerve implementation in ASCENT and Sim4Life.

Multifascicular dummy nerve model

We validated activation thresholds for fibers seeded in a multifascicular dummy nerve instrumented with a bipolar cuff electrode (Figure D and E). The segmented histology was created using our mock_morphology_generator.py script (Mock Morphology).

_images/Picture5.jpg

Figure D. Mock morphology inputs to the define tissue boundaries for a multifascicular dummy nerve for validation with Sim4Life. Scale bar is 100 µm long. The nerve is a perfect circle (diameter = 250 µm, centered at (x,y)=(0,0) µm). The inners are also perfect circles: (1) diameter = 50 µm, centered at (x,y)=(40,50) µm, (2) diameter = 60 µm, centered at (x,y)=(-50,0) µm, and (3) diameter = 80 µm, centered at (x,y)=(20,-60) µm.

_images/Picture6.jpg

Figure E. FEM of a multifascicular nerve sample instrumented with a bipolar cuff electrode.

The conductivity values applied to multifascicular nerve sample finite element model are provided in Table A, and the boundary conditions applied are provided in Table B.

We seeded a single 5.7 µm diameter fiber in the center of each fascicle. Between the ASCENT and IT’IS implementations, there was less than a 3% difference in threshold to a single 100 µs duration monophasic rectangular pulse.

Multifascicular human nerve model

We validated activation thresholds for fibers seeded in a multifascicular human cervical vagus nerve instrumented with a LivaNova bipolar cuff electrode (Figure F and G). The segmented histology was created using Nikon NIS-Elements.

_images/Picture7.jpg

Figure F. Raw histology image (r.tif), segmented inners (i.tif), segmented nerve (n.tif), and scale bar (s.tif) of a human cervical vagus nerve sample that served as inputs to define the cross-section of the nerve in the FEM for validation with Sim4Life.

_images/Picture8.jpg

Figure G. FEM of a human cervical vagus nerve sample instrumented with a LivaNova cuff electrode.

The conductivity values applied to the human cervical vagus nerve sample finite element model are provided in Table A, and the boundary conditions applied are provided in Table B.

We seeded 5.7 µm diameter fibers in each fascicle. Between the ASCENT and IT’IS implementations, there was less than 2.5% difference to a single 100 µs duration monophasic rectangular pulse.

Comparison of MRG fit to Bucksot 2019

Comparison of MRG fit to Bucksot et al. 2019

_images/Picture11.jpg

Figure H. Our piecewise polynomial fits to published MRG fiber parameters compared to the Bucksot et al. 2019’s interpolation [Bucksot et al., 2019]. Single quadratic fits were used for all parameters except for internode length, which has a linear fit below 5.643 µm (using MRG data at 2 and 5.7 µm) and a single quadratic fit at diameters greater than or equal to 5.643 µm (using MRG data >= 5.7 µm); 5.643 µm is the fiber diameter at which the linear and quadratic fits intersected. The fiber diameter is the diameter of the myelin. “Paranode 1” is the MYSA section, “paranode 2” is the FLUT section, and “internode” is the STIN section. The axon diameter is the same for the node of Ranvier and MYSA (“node diameter”), as well as for the FLUT and STIN (“axon diameter”). The node and MYSA lengths are fixed at 1 and 3 μm, respectively, for all fiber diameters.

Micro Leads cuff measurements

We collected and measured images of 200, 300, and 400 µm Micro-Leads Neuro cuffs (Somerville, MA) (Figure I and Table C).

_images/Picture17.jpg

Figure I. Micro-Leads cuff measurements taken for 200, 300, and 400 µm inner diameter cuffs. Inner diameter (ID) displacement is defined from the center of the circle of the outer diameter (OD) to the center of the semicircle for the inner diameter.

Table C. Cuff measurements (units: micrometer) for 200, 300, and 400 µm inner diameter Micro-Leads cuffs.

Cross-Sectional Measurements Longitudinal Measurements
Cuff # cuff measured OD ID ID Displacement (x) Peak (x) Tangent (x) Gap (y) End Insulation Contact Width Contact Spacing
200 μm 6 bipolar 1080 210 135 15 175 30 295 590 305
300 μm 12 bipolar, 2 tripolar 1190 265 130 30 255 40 415 680 390
400 μm 4 bipolar 1605 465 325 0 200 35 535 745 475

We also collected and measured images of 100 µm Micro-Leads cuffs, which had a different cross-section from the larger diameter cuffs and are therefore reported separately (Figure J and Table D).

_images/Picture18.jpg

Figure J. Micro-Leads cuff measurements taken for 100 µm inner diameter cuffs. Inner diameter (ID) displacement is defined from the center of the circle of the outer diameter (OD) to the center of the semicircle for the inner diameter.

Table D. Cuff measurements (units: micrometer) for 100 µm inner diameter Micro-Leads cuffs.

Cross-Sectional Measurements Longitudinal Measurements
Cuff # cuff measured OD ID Trapezoid Base Tangent (x) Gap (y) ID Displacement (x) End Insulation Contact Width Contact Spacing
200 μm 6 1060 125 200 85 30 210 180 320 150

For all “preset” cuff JSON files based on our Micro-Leads cuff measurements, we recessed the contacts by 50 µm. Though we were not able to directly measure the recess depth of the contacts, Micro-Leads informed us that the surface of the electrode metal could never be recessed by more than ~60-70 µm.