ASCENT Validation
Sim4Life validation
We designed test simulations to verify ASCENT’s activation thresholds. The verifications were performed by The Foundation for Research Technologies in Society (IT’IS) with the Sim4Life (Zurich, Switzerland) simulation platform. Running the following simulations required modification of the Sim4Life solver to implement the required electrical anisotropy of tissue conductivities and the boundary condition to represent the thin layer approximation used to model the perineurium.
Monofascicular rat nerve model
We validated activation thresholds for fibers seeded in a model of a rat cervical vagus nerve instrumented with a bipolar cuff electrode (Figure A and B).
The conductivity values applied to the rat cervical FEM are provided in Table A, and the boundary conditions applied are provided in Table B.
Table A. Conductivity values for FEM of rat cervical vagus nerve. These values were also used in multifascicular nerve model and human model verifications.
Parameter | Application | Resistivity Ω-m |
---|---|---|
Endoneurium | Within each fascicle | 1.75 longitudinal 6 radial (rat and human) 12 radial (multifascicular dummy model) |
Saline | Cylindrical shell between the nerve and cuff | 1/1.76 |
Platinum | For both contacts | 1/(9.43*106) |
Silicone | For electrode body | 1012 |
Muscle | Used for the surrounding "medium": Everything outside of the nerve and cuff, other than the saline layer between the nerve and cuff | 1/0.35 longitudinal 1/0.086 radial |
Epineurium | Within the nerve around each fascicle | 6.3 |
Encapsulation | Between cuff and nerve, and immediately surrounding cuff | 6.3 |
Table B. Boundary conditions used in FEM of rat cervical vagus nerve, multifascicular dummy nerve, and human cervical vagus nerve.
Parameter | Setting |
---|---|
Current conservation | All domains |
Initial condition | V=0 (all domains) |
Perineurium (sides of each fascicle) | Modeled as contact impedance 1149 Ω-m *0.03* dfasc[m] |
Ground (all outer boundaries of the model) | V = 0 |
Point current source (one in each contact) | -1 mA 1 mA |
We compared thresholds for 100 5.7 µm myelinated axons (MRG model) seeded in the cross-section of the nerve in response to a single 100 µs duration monophasic rectangular pulse. The differences in thresholds between ASCENT and IT’IS model implementations was <4.2% for all fibers, demonstrating strong agreement (Figure C).
Multifascicular dummy nerve model
We validated activation thresholds for fibers seeded in a
multifascicular dummy nerve instrumented with a bipolar cuff electrode
(Figure D and E). The segmented histology was created using
our mock_morphology_generator.py
script (Mock Morphology).
The conductivity values applied to multifascicular nerve sample finite element model are provided in Table A, and the boundary conditions applied are provided in Table B.
We seeded a single 5.7 µm diameter fiber in the center of each fascicle. Between the ASCENT and IT’IS implementations, there was less than a 3% difference in threshold to a single 100 µs duration monophasic rectangular pulse.
Multifascicular human nerve model
We validated activation thresholds for fibers seeded in a multifascicular human cervical vagus nerve instrumented with a LivaNova bipolar cuff electrode (Figure F and G). The segmented histology was created using Nikon NIS-Elements.
The conductivity values applied to the human cervical vagus nerve sample finite element model are provided in Table A, and the boundary conditions applied are provided in Table B.
We seeded 5.7 µm diameter fibers in each fascicle. Between the ASCENT and IT’IS implementations, there was less than 2.5% difference to a single 100 µs duration monophasic rectangular pulse.
Comparison of MRG fit to Bucksot 2019
Comparison of MRG fit to Bucksot et al. 2019
Micro Leads cuff measurements
We collected and measured images of 200, 300, and 400 µm Micro-Leads Neuro cuffs (Somerville, MA) (Figure I and Table C).
Table C. Cuff measurements (units: micrometer) for 200, 300, and 400 µm inner diameter Micro-Leads cuffs.
Cross-Sectional Measurements | Longitudinal Measurements | |||||||||
---|---|---|---|---|---|---|---|---|---|---|
Cuff | # cuff measured | OD | ID | ID Displacement (x) | Peak (x) | Tangent (x) | Gap (y) | End Insulation | Contact Width | Contact Spacing |
200 μm | 6 bipolar | 1080 | 210 | 135 | 15 | 175 | 30 | 295 | 590 | 305 |
300 μm | 12 bipolar, 2 tripolar | 1190 | 265 | 130 | 30 | 255 | 40 | 415 | 680 | 390 |
400 μm | 4 bipolar | 1605 | 465 | 325 | 0 | 200 | 35 | 535 | 745 | 475 |
We also collected and measured images of 100 µm Micro-Leads cuffs, which had a different cross-section from the larger diameter cuffs and are therefore reported separately (Figure J and Table D).
Table D. Cuff measurements (units: micrometer) for 100 µm inner diameter Micro-Leads cuffs.
Cross-Sectional Measurements | Longitudinal Measurements | |||||||||
---|---|---|---|---|---|---|---|---|---|---|
Cuff | # cuff measured | OD | ID | Trapezoid Base | Tangent (x) | Gap (y) | ID Displacement (x) | End Insulation | Contact Width | Contact Spacing |
200 μm | 6 | 1060 | 125 | 200 | 85 | 30 | 210 | 180 | 320 | 150 |
For all “preset” cuff JSON files based on our Micro-Leads cuff measurements, we recessed the contacts by 50 µm. Though we were not able to directly measure the recess depth of the contacts, Micro-Leads informed us that the surface of the electrode metal could never be recessed by more than ~60-70 µm.